The Calibration Control Kit HandbookChapter 13 of 22

Records that hold up

An auditor tests a calibration system by following a chain: from a part's inspection record to the instrument, from the instrument to its certificate, from an out-of-tolerance certificate to its assessment, and from a key characteristic to its gauge study. Every link must be a record you can pull in minutes.

RecordWhat it must showKept where
Equipment registerEvery instrument, use, risk, interval, status, last known goodE01
CertificatesAs-found and as-left data, uncertainty, traceability, standards used, conditionsQuality drive by instrument ID
Inspection recordsThe instrument ID used for each characteristicTravelers, inspection reports
Interim check recordsDate, check, result, initialsAt the station; scanned monthly
Impact assessments and the OOT logWindow, trace, evaluation, decision, notification, closureE02
MSA plan and studiesStudy, result, verdict, conditions, next dueE03, E04
Interval changesThe old and new interval and the reasonE01 notes
Laboratory approvalsAccreditation and scope; approval on the supplier listSupplier file

The findings we see most, and how the kit prevents them

FindingUsual root of itWhere the kit prevents it
Out-of-tolerance result with no product assessmentNo trigger between the certificate and the assessmentCertificate checklist (chapter 6); E02 opened the same day
Instrument in use past its due dateRecall not reviewed; shutdown not plannedE01 status and Summary; weekly review
Instrument on the floor not on the registerEmployee-owned or new tools not enteredFloor walk; register scope (chapter 5)
Inspection records do not identify the instrumentRecord designRequired instrument ID field; quarterly trace test
Certificate shows no as-found data or uncertaintyPurchase order does not ask for themLab requirements on the order (chapter 6)
No evidence a gauge suits a key characteristicCalibration taken as proof of fitnessE03 plan; E04 study
Capability claimed on an unproven gaugeStudy order reversedE06 setup names the gauge study
CMM program changed without verificationNo change control on programsChapter 12; MSA plan line
Intervals never reviewedNo use of as-found historyChapter 7; monthly review
Customer not told about shipped product in questionWaiting for a complete assessmentE02 Summary flags notification; chapter 9

Need more than one resource, or not sure where to start? Book a complimentary 30-minute Aegis Quality Guidance Call. We'll find the tools that fit your situation and a sensible place to start. No sales pressure, no obligation.

A management aid, not certification or a registrar audit. Using it does not by itself show conformity to AS9100D or any other standard, and your customers' requirements may add to it.

Clause numbers refer to AS9100D (SAE AS9100 Rev D) unless another document is named. The standards' text is not reproduced here: use your own licensed copy.

Aegis Insights are our own guidance, labeled wherever they appear: how we approach this work, not a requirement of any standard or customer.