Every measurement has some doubt in it: the instrument's own uncertainty, the reference standard's, temperature, the part's form, the operator. It matters most when a reading lands close to a limit.
- Accuracy ratio. A common rule of thumb is that the instrument's uncertainty should be no more than a quarter of the tolerance it checks (a 4 to 1 ratio), and many shops aim for a tenth. When the ratio is poor, a reading near the limit cannot tell a good part from a bad one.
- Decision rules. Decide in advance what happens to a reading close to a limit: accept it as read, apply a guard band (accept only inside a narrower limit), or re-measure with a better instrument. Write the rule into the inspection instruction, and follow your customer's rule where it sets one.
- Certificates. Read the lab's stated uncertainty against the tolerances the instrument checks. A lab uncertainty that is a large share of a tight tolerance is a reason to change the instrument or the lab.