The Calibration Control Kit HandbookChapter 7 of 22

Setting and adjusting intervals

An interval is a bet on how long an instrument stays within its limits in your conditions of use. Start from a sensible default, then let the as-found history move it. Write the reason for every interval change on the register.

Kind of instrumentStarting interval we often useShorten whenLengthen when
Calipers, micrometers in daily use6 to 12 monthsAny out-of-tolerance result; heavy use on hard or abrasive materialsSeveral calibrations in a row with as-found results well inside the limits
Bore, plug, ring and thread gauges6 to 12 months (by use count where heavy)The go member trending toward its wear limitLight use and stable wear readings
Height gauges and indicators6 to 12 monthsDrift in the as-found data, even in toleranceStable history and a daily check on a block
CMM and vision systems12 months, with probe qualification at each start-upCrashes, moves, repairs, temperature problemsRarely: the interim artifact checks carry the confidence
Torque tools6 to 12 months or by cycle countUse near the top of the range; any dropLight use and stable results
Gauge blocks, masters, surface plates12 to 36 monthsDamage, heavy use as a setting masterLab use only and stable history

Two rules keep intervals honest. First, any out-of-tolerance result shortens the interval until the history says otherwise. Second, a run of results that drift toward a limit, while still inside it, is a reason to shorten before it fails, not after.

Adjusting an interval: an example

Three calibrations of a height gauge, all in tolerance, against an accuracy limit of plus or minus 0.020 mm. The as-found history is the evidence for the decision; the register notes carry the reason.

CalibrationLargest as-found errorShare of the limitWhat we would decide
Year 1+0.004 mm20%Keep 12 months
Year 2+0.009 mm45%Keep 12 months, but the drift is now visible: add a monthly block check
Year 3+0.015 mm75%Shorten to 6 months: on this trend the next one fails
Two 6-month results under 25%+0.004 mm, +0.005 mm20%, 25%Return to 12 months, keep the block check, write the reason

Out-of-tolerance results move the interval at once. In-tolerance results move it on a pattern: one good result says little, a run of them says a lot. A written rule of thumb that the whole team uses is better than a clever method nobody follows.

Need more than one resource, or not sure where to start? Book a complimentary 30-minute Aegis Quality Guidance Call. We'll find the tools that fit your situation and a sensible place to start. No sales pressure, no obligation.

A management aid, not certification or a registrar audit. Using it does not by itself show conformity to AS9100D or any other standard, and your customers' requirements may add to it.

Clause numbers refer to AS9100D (SAE AS9100 Rev D) unless another document is named. The standards' text is not reproduced here: use your own licensed copy.

Aegis Insights are our own guidance, labeled wherever they appear: how we approach this work, not a requirement of any standard or customer.